Calibration
Calibrate your Data Walker against a known distance for maximum measurement accuracy.
When to Calibrate
- After initial setup or reflashing firmware
- When moving the device to a different walker
- If measured distances consistently differ from expected
- After replacing wheels or magnets
- Periodically (recommended: monthly) for quality assurance
Calibration Course Setup
For calibration, you need a pre-measured 3-meter distance:
- Mark a start line and end line exactly 3 meters apart
- Use a calibrated measuring tape or wheel for accuracy
- Ensure the floor is flat and similar to clinical testing conditions
- The path should be clear in both directions for forward and reverse walking
Running Calibration
- From the main menu, select “Calibration”
- Position the walker at the start line
- Press the button to begin recording the forward phase
- Walk forward at a normal pace to the 3-meter end line
- Stop at the end line (do not turn around)
- Press the button to end the forward phase
- Press the button again to begin recording the reverse phase
- Walk backward in reverse, retracing the same 3-meter path to the start line
- Press the button to end the reverse phase and complete calibration
- The device will compare the measured distance to the expected 6 meters (3m forward + 3m reverse)
Interpreting Results
After completing the calibration walk, the firmware calculates the measured distance and compares it to the expected 6 meters. If there is drift, you will see:
Measured: 5.92 m
Expected: 6.00 m
Suggested factor: 1.0135
Apply calibration? [Yes/No]
If the measured distance is close to expected (within ±0.5%), no calibration adjustment will be suggested.
Applying Calibration
Select “Yes” to apply the suggested calibration factor. The new factor is stored in non-volatile memory and persists across power cycles and firmware updates.
Manual Calibration
To manually set a calibration factor:
- Navigate to Settings → Calibration
- Select “Manual Entry”
- Enter the calibration factor (0.9000 – 1.1000)
- Confirm to save
Resetting Calibration
To reset to the default calibration:
- Navigate to Settings → Calibration
- Select “Reset to Default”
- Confirm to restore factor to 1.0000
Troubleshooting
Calibration factor out of range
If the suggested factor is outside 0.9000–1.1000, there may be a hardware issue. Check:
- Wheel diameter matches expected specifications
- All 3 magnets are present and detected
- Sensor is properly positioned
Inconsistent measurements between runs
High variance between runs may indicate:
- Loose magnet(s) — run self-test to verify
- Wheel slippage on smooth floors
- Inconsistent walking path