Power-On Self-Test
The self-test runs automatically after flashing or power-on, verifying the Hall sensor and magnet configuration.
What It Tests
The self-test verifies:
- Magnet detection: All 3 magnets are detected during one wheel rotation
- Even spacing: Magnets are approximately 120° apart (within tolerance)
- Signal quality: Clean transitions without noise or bounce
- Sensor connectivity: RJ22 cable is properly connected
Running the Self-Test
- Power on the device or flash new firmware
- The display shows “Self-Test” with instructions
- Slowly rotate the wheel one complete turn (360°)
- Maintain a steady pace — not too fast, not too slow
- Wait for the result to display
Self-Test Passed
When successful, you will see “Self-test passed” and the device proceeds to the main menu. The calibration factor defaults to 1.0000.
Self-Test Failures
“Only X magnets detected”
The test detected fewer than 3 magnets. Possible causes:
- A magnet has come loose or fallen off
- Sensor is too far from the magnet path (should be within 5mm)
- You did not complete a full rotation
- Wheel was rotated too fast
“Uneven magnet spacing”
The magnets are not evenly distributed. This affects measurement accuracy.
- Verify magnets are positioned 120° apart
- Check that no magnets have shifted
- Re-mount magnets if spacing is incorrect
“No signal detected”
The Hall sensor is not sending any data.
- Check the RJ22 cable connection at both ends
- Verify the cable is not damaged
- Ensure the sensor is receiving power (3.3V)
- Test with a multimeter if available
Bypassing the Self-Test
Not Recommended
The self-test exists to ensure measurement accuracy. Bypassing it may result in unreliable clinical data. Only bypass for troubleshooting purposes.
To bypass the self-test, hold both buttons during power-on until the main menu appears. This is intended only for diagnostics.
Re-Running the Self-Test
To run the self-test again without reflashing:
- From the main menu, navigate to Settings
- Select “Run Self-Test”
- Follow the on-screen instructions
Next Steps
After passing the self-test, proceed to calibration for maximum accuracy, or start using the device with the default calibration.